Utility App for wireless control of cold behaviour analysers from ORBIS BV that are used in Petroleum & Petrochemical Laboratories.
The app is capable of controlling the cold behaviour analysers of ORBIS BV, provides information about the analysers and reports the results of the analysers in a tabular and graphical manner.
A set of standard methods allows to easily set up measurement programs that can be send to one of the connected analysers.
A status screen provides information about the current status of each sample currently running on one of the analysers.
Results can be viewed real time while the instrument is running or upon completion of the test.
Results can be exported via ftp connection in xml format to a remote database (LIMS) Laboratory information management system. or printed on a Zebra thermal printer.
The format of reporting to the printer or LIMS can be set easily on a formatting screen which can be set as a global default setting or for each standard method or program individually.
Quoi de neuf dans la dernière version ?
- Updated to latest Xcode version
- fixed a problem where the app did not save the database
- fixed a problem where the bath settings in the program could not be changed
- fixed a problem where the aspiration time of the last aspiration where not updated in the state view.
- fixed a problem where new programs should not be changed or edited or deleted
- fixed a type error in the message for Sensitivity warning PP
- fixed a problem of not showing bath calibration date on report
includes new FW:
Bath v2.3.4
Introduced networking fixes to increase stability and responsiveness of bath over the network.
Reworked DCP leasing to prevent duplicating hostnames, MACs and IPs.
Fixed memory leak in telnet when a connection fails (closes socket before reopening).
Close and reopen socket on bath and head disconnects.
Fixed overrun issue due to HTTP requests
Fixed issue with init order of ethernet and network stack, where network stack was initialised before ethernet was ready
Fixes issue with TCB resets not happening on connect timeouts
Does not clean client list, but tracks IP's based on slot numbers without any sorting, providing a more stable method of distributing IP's. No more conflicts.
Able to correctly store what IP's were declined so no more repeat cycles of trying to request the same IP over and over again.
Always updates when uploading FW, no more date checking
CFPP v1.7.5
Fix bot parm default setting for only new devices with LB version 2. Set to 700 instead of 1400.
Fixed a critical safety issue where a program fault could leave the pump at full speed until watchdog reset. The pump is now
immediately disabled on entry to a stuck HardFault state.
CPPP v1.17.1
Allow negative cloud point measurements (cp_a) to trigger detection
Read cal mem of pt100 only once on connect
add magic check on parms load
Rework of pairing firmware and wifi scanning for more reliably connectivity
When the CPPP head loses WiFi or reboots it will try to connect to the configured SSID 3 consecutive times on 5 second intervals. When it doesn't find a network it will either:
Powered: go into docking station mode
Unpowered: turn off after 30 seconds
To reconnect CPPP the user has to change the power input, for example:
- Lifting it from the docking station, when the docking station is powered
- Turning on the ColdBlock with the head seated
- Shortly lifting and seating the head from the ColdBlock
To summarise: anything that changes the power input state of the head will restart the connectivity.
Docking station mode is entered when the head does not see a response to the blinking sequence on the WiPower connection. It wil assume that the bath is not registering the blinking sequence and thus will conclude it is seated on a docking station.
Fix for issue when CPPP head would not connect after ColdBLock was powered on.
Fix for docking station detection, where CPPP sometimes thought it was on a docking station when it was not.
CPPP STEP v1.0.1
Vscode config added.
Build system updated.
Drv8434 motor driver implemented.
Hw version detection added, the code is compatible with older devices.